0.3 The LeicaTCS-NT Confocal Microscope |
The Microscope:
The LeicaTCS-NT laser scanning confocal microscope is built around a Leica DMRI inverted microscope, which features
electronic focus and objective changer controls. A manual X-Y mechanical stage holds
standard slides. A stage adapter and chamber may be used to accommodate 18 mm round
cover slips for cultured cells or a heated stage for 25 mm round cover slips may be
attached. For best optical performance number 1.5 cover slips (170 um thickness)
should be used. The step button on the front panel
of the scope changes the fine focus increment size, S1 being the finest step and S3
the courses.

| Lens | Mag | NA | Imm. | Working | Zoom for max. resolution | Approx. Optical thickness pin hole 1.0 Airy disc unit |
Measured |
| Fluor | 10X | 0.30 | air | 4.90 mm | ~1 | 40 um approx. | 11. um |
| Plan Apo | 20X | 0.70 | oil/gly/H2O | 250 mm | ~2 | 5 um approx. | 5.3 um |
| Plan Apo | 40X | 1.25 | oil | 170 um | 4.0 | 1.0 um approx. | 1.2 um |
| Plan Apo | 100X | 1.40 | oil | 90 um | 2.5 | 0.6 um approx. | 0.4 um |
Note that the 10X objective is suitable for sample location and orientation. The 20X, 40X and 100X objectives should be used with immersion oil. Differential Interference Contrast (Nomarski) optics is available with the 40X and 100X objectives.
Binocular viewing:
Standard wide field viewing. Dichroic filters for
Fluorescence (FITC), GFP, YFP (filter wheel position 3) and Rhodamine (TRITC, Texas Red)
(filter wheel position 2) with a 50 Watt Mercury Lamp for excitation allow wide field appraisal of fluorescence and
convenient location of desired field of view for confocal scanning. DIC
(Differential Interference Contrast) can be used while viewing through the oculars or
scanning confocally.
Confocal scanning:
Confocal scanning can be carried out with up to three fluorescence channels and one
transmitted light channel in any combination. Filters and lasers are set up for
viewing of FITC (Fluorescence) like, Rhodamine like (or TRITC or Texas Red) and CY5 like
fluorophores in either simultaneous scanning or sequential scanning. Transmitted
light detection can be used either with laser scanning wide field (like standard
transmitted light microscopy) or Differential Interference Contrast (DIC, very similar to
Nomarsky).
Excitation Lasers:-
Three individual lasers are used. Argon 488 nm (blue excitation for FITC or CY2 and
similar fluorophores) > 25 mW & 476 nm <5mW. Krypton 568 (green excitation
for Rhodamine or Texas red and similar fluorophores) > 25 mW, Neon-Helium 688 nm (red
excitation for CY5 and similar fluorophores) 25 mW. Laser power transmitted from
each laser is controlled by an acusto optical tunable filter (AOTF), which mixes the proportion of light from each laser
transmitted to the specimen. Many fluorochromes may be accommodated
with these three lasers. Please note that the emission spectra graph is
intended as a rough guide of the laser characteristics and does not reflect the actual
response which is filtered through the AOTF and will also vary with laser manufacturer and
model. Confocal detector pinhole size, zoom factor, detector sensitivity and
focus are controllable during scanning via the seven knob panel box or mouse
controls. Transmitted laser light images can be collected from the laser light
passing through the sample while confocal scanning is being performed.
The Acoustical Optical Tunable Filter (AOTF):-
The AOTF mixes the laser light emitted from the three lasers. The settings shown
below are roughly optimum initial settings on our confocal scope. The values may
need to be adjusted for example to reduce photo-bleaching. Currently the 488 nm
laser is operating efficiently and 5% may be a more appropriate power level. These
percentages represent the fraction of power emitted from each laser. The Argon and
Krypton lasers have front panel power controls, which vary the power presented to the
AOTF. Therefore power can be altered without changing the setting on the
AOTF. In other words the AOTF settings represent the percentage of power emitted
from the lasers, which is being transmitted to the scanhead.

Pinhole, Detector & Filters:-
The confocal scanhead has the three PMTs (photo multiplier
tubes) for the detection of fluorescence. The detector pinhole, which is a key
element of a confocal system, lies before the PMTs and filters. The pinhole size is
adjustable and varies the amount of out of focus light reaching the PMT detectors.
Various dichroic mirrors (wave length selective mirrors) and plain mirrors may be selected
to direct the emitted light from the specimen to the PMT detectors. Barrier filters
may be placed in front of the PMT detectors in order to remove reflected light from the
specimen and improve the wavelength selectivity of the detected light. Also laser
light passing through the specimen can be collected by the condenser, which focuses this
light onto a fourth PMT detector, which forms a transmitted light image at the same time
as a fluorescence image is gathered confocally. In effect the condenser is being
used as an objective! Conveniently, the mirrors and dichroics along with default
laser power and PMT settings may be automatically selected for common fluorophores using
preset configurations listed in the Filter Settings box. If required they may
be customized manually.
LP = long pass
BP = band pass
RSP = reflection short pass
DD = double dichroic
TD = triple dichroic
N.B.
long = long wavelengths,
short = short wavelengths
Image Storage and Transfer:
The confocal control and image processing computer is a WindowsNT PC 200 MHz Pentium
Pro workstation, which is networked to the campus via TCP/IP, NetBEUI (Microsoft Windows
Network), Apple Talk, Novel IPX network protocols. Single images are stored as TIFF
files and multi-channel fluorescence or z-series images are stored as multipage
TIFF files. On the computer, image storage is in users' sub-directories located
at D:\USERS1 and D:\USERS2 and D:\USERS3 and
I:\ and J:\ and optionally M:\
(for Apple Talk access). Files may be transferred and stored directly to Zip or Jaz
disks.
Image Processing:
Some image processing may be performed with the confocal acquisition program
(LeicaTCS-NT). Single channel images are viewable and editable with PhotoShop (Mac
or PC). Multi channel tiffs may by viewed with WangImage or Kodak Image (part of
Windows95/98/NT) or with ThumbsPlus. To view multi-channel tiff files they can be
split into individual tiff images using tiffsplit4.exe (PC only). Some software is available for bulk
processing of images and 3-D rendering, viewing, production of computer movies of time
lapse events, rotations, 3D-stereo pairs, etc.
|
|
|
|
|
© 1998-2003 Michael Chua, Cell & Molecular Physiology, UNC. All rights reserved. | |
| To Confocal Homepage | To Operating the confocal | To Booking the confocal | Go back a page | Comments: Dr. Michael Chua |