Most user-caused errors and inefficiencies are discussed in Smooth Starts, and one of
those is recalled here. The rest of this list includes several ways in which things go
wrong or just hang up for no apparent reason.
- Can see specimen through oculars but no image is seen on the computer display
panel while scanning.
Check for the presence of scattered laser light from specimen on the stage.
If none ensure that scan head shutter rod is pulled out. Give it a firm tug.
Check that laser power is not at zero or too low. Check sensitivity (PMT
voltage). Check focus (see 2). Open pinhole to at least 1.0
- Sudden loss of focus in the conventional microscope or when initially scanning.
The stage position of the microscope is really quite stable. Sudden loss of focus
is almost always due to one of two things: 1) an arm brushing the focus knob when reaching
over to the sliders or fluorescent filters and 2) establishing initial focus on the sample
while the galvo stage is in the parked position. The galvo stage is in this lowered
position when the TCS program is not loaded or the scan electronics are off. Upon starting
of the TCS program, the stage rises by about 85 m m (one-half
of its total throw of 170 m m) to its operational and holding
position. Focus is lost, of course. Starting from the original focus position, the new
focus position can usually be found quite simply by raising the lenses 85 m m, as indicated by the LCD. See also Smooth Starts.
- Focusing is at higher level than previous positions.
This can happen when slide is being hinged up by the objective pushing on
the slide. One slide of the slide remains on the stage while the spring clips allow
the other edge to lift up off the stage. A wedge profile of oil forms between the
objective and the slide creating a gap equal to the working distance of the
objective. Typically the objective is about 400 mm
higher than the upper limit, if the limit has been set correctly. This situation is
a disaster for z-series sections, and can be a bigger disaster if another objective is
swung into place, causing the nose of the objective slide to scrape into the edge of the
slide. Solution: move objective close to slide and find focal plane while
increasing distance between the slide and the objective. Typically z-position on
microscope should not be more than 200